Main Features
1.Intelligent: Adopts high-precision AD conversion and single chip microcomputer processing technologies.
2.It can be used for the measurement of the NaOH concentration and temperature, automatic temperature compensation, automatic range switching and self-checking.
3.Reliability: All the components are arranged on one circuit board. No complicated functional switch, adjustingknob or potentiometer is arranged on this instrument.
4.Strong interference immunity: Its current output adopts optoelectronic isolating technology.
5.This meter has strong interference immunity and the capacity of long-distance transmission as well as excellent electromagnetic compatibility.
6.Waterproof and dust-proof design: Its protection grade is IP65. It is applicable for outdoor use.
7.Automatic range switching: The measuring ranges can automatically switch within the measuring range of the
electrode.
8.Communication interface: It can be easily connected to a computer to perform monitoring and communication.
Phase sensitive detection: Eliminating the influence of conductors on the measurement of hydrochloric acid
concentration.
9.Automatic measuring frequency switching: This can help prevent electrodes from polarization, so as to enhance
their accuracy.
Main parameters
Measuring range and error | (0~1.0)%NaOH; (0~30)%NaOH; temperature: 0~99.9C |
Intrinsic error of electronic units | Concentration: ±1.0%, temperature: ±0.3C |
Intrinsic error of the instrument |
Concentration: ±2.0% (the error may be less than 0.05% after usual point calibration) Temperature:±0.5C(0.0C≤T≤80.0C) |
Repeatability | ±1.0% |
Stability | ±1.0%/24h |
Automatic temperature compensation range | 0~99.9C |
Solution tested | 0~99.9C, 0.3MPa |
Output current error | ≤±1%FS |
Communication interface | RS485 Communication Interface |
Protection | lP65 |
Isolated current output | 4~20 mA(load <750Ω) |
Compensation range of the temperature coefficient | (0~10.0)%/C(adjustable) |
Alarm relay | AC220V, 3A |
Data storage capacity | l month (1 point/5 minutes) |
Saving time of data under a continuous power-failure condition | 10 years |
Power supply | AC220V±22V, 50Hz±1Hz |
Overall dimension of the instrument | 146 (length) x 146 (width) x 108 (depth) mm |
Dimension of the hole | 138 x 138mm |
Working conditions | Ambient temperature: 0-60C; relative humidity <85% |
Clock accuracy | ±1 minute/month |